Could you provide an overview of Arbin battery holders and racks?

Von Arbin Team | Juni 19, 2026

In general, we offer holders for cylindrical, pouch, coin, and prismatic cells. We also provide battery racks for both coin cells and cylindrical cells. Cell type Max current Coin cell 3 A, 5 A Pouch cell 10 A, 60 A, 300 A Cylindrical cell 10 A, 30 A, 60 A, […]

Can Arbin MZTC integrate with other companies’ cyclers?

Von Arbin Team | Juni 19, 2026

Physically, yes. The MZTC can be connected to the cycler through the charge/discharge port. On the MZTC, each mini chamber’s temperature can be controlled directly using the built-in LCD screen. However, we currently do not have a communication protocol that allows the cycler to control the MZTC through software.

Can the Arbin MZTC control temperatures below 0°C? If not, what is its temperature range?

Von Arbin Team | Juni 19, 2026

No. The MZTC cannot control temperatures below 0°C. The widest temperature range of the MZTC is 10°C to 60°C. However, the actual operating range may be narrower depending on the ambient temperature: The lower limit is 10°C or ambient temperature minus 10°C, whichever is higher. The upper limit is 60°C […]

Can I test a battery without convoluted wiring connections?

Von Arbin Team | Juni 15, 2026

Yes. Arbin offers several products that can help simplify and organize the connection between the battery and the cycler. I. Single Coin Cell Holder With the Phoenix connector, the holder can be plugged directly into the machine’s channel port. II. 8-Channel Universal Coin Cell Battery Holder Tray Part Number: 409716 […]

Gibt es spezielle Anforderungen an die Vorbereitung der Einrichtung für die Installation des Arbin-Systems?

Von Arbin Team | Mai 8, 2026

Generally, there are several points to note: Arbin machines typically have a working ambient temperature range of 0°C to 30°C. Therefore, please prepare the air-conditioning system accordingly. The power grid system must be prepared to match the machine’s input power requirements. For example, some machines require single-phase power input, while […]

Kann ich das Arbin-System selbst für Hardware und Software einrichten oder benötige ich die Unterstützung von Arbin?

Von Arbin Team | Mai 8, 2026

Arbin systems will be shipped with a computer that includes the full software kit pre-installed. The system hardware setup is simple: customers can easily set it up themselves by connecting the cycler to the computer and auxiliary devices using Ethernet cables. Figure 1: Example TCP/IP Connection to Arbin Main System […]

Welche Zahlungsmethoden werden beim Kauf einer Arbin-Maschine akzeptiert?

Von Arbin Team | Mai 8, 2026

Normal bank wire transfer is preferred. We avoid LC. Arbin’s standard payment mode is 40% – 40% – 20%. The first 40% is due right after the PO is received. Production starts only after Arbin’s bank account receives this payment. Lead time counts from this moment, not from PO receipt. […]

”Conflict Clamp V“ erscheint, wenn ich einen Test starte

Von Arbin Team | Mai 8, 2026

This error occurs only when the Channel Based Voltage Clamp option is unchecked in ArbinAdvSys.cfg. This means all IV channels on the same IV channel board share one voltage clamp/voltage check. The MITS software must ensure all schedules on the same IV board channels share the same Voltage Clamp setting; […]

Mein Test wurde aufgrund eines Fehlers namens “Aux CAN Lost” oder ”Aux ACK Failed” abgebrochen.”

Von Arbin Team | Mai 8, 2026

This error means the Aux channel lost communication with IV channels. It usually occurs when the Batch/Mapping file was created or modified without downloading it to the cycler. Downloading the Batch/Mapping file resolves this issue: If this does not resolve the issue, check the cable connection from the Aux module […]

Müssen die Tests in einer Temperaturkammer, einer Sicherheitsschrank oder einer Schutzgasbox stattfinden?

Von Arbin Team | Mai 8, 2026

Installing a battery inside the chambers or glovebox is optional, Arbin can integrate it if customers need it for testing. Arbin offers an 8-zone temperature chamber for battery testing (MZTC) with an operating temperature of 10-60C and for cells rated up to 60A. Connections are made easy with the cables […]

Wie wird das zu testende Gerät (Zelle, Modul oder Pack) mit dem Arbin-System verbunden?

Von Arbin Team | Mai 8, 2026

There are several way to connect to the DUT as follows: Individual cables Individual cell holders (coin cells, cylindrical cells, flat/pouch cells) Trays for holding multiple cells (coin cells, cylindrical cells, flat/pouch cells) Racks for organizing cells Chambers We will go to each method one by one as follows: 1. […]

Meine Prüfung brach wegen eines Fehlers mit der Bezeichnung “IR Inf” ab.”

Von Arbin Team | Mai 6, 2026

The “IR Inf” error occurs when the system fails to obtain a valid result during an Internal Resistance (IR) measurement step. This typically happens when the measurement conditions are not sufficient to produce a reliable calculation, causing the system to return an infinite (invalid) resistance value and stop the test. […]

Kann ich eine MTCI-Einheit kaufen und zwischen den Kammern wechseln?

Von Arbin Team | Mai 6, 2026

The answer is yes and no. Yes, if those chambers use the same communication protocol. For example, Votsch VT4002 and Weiss WT20/40 use the Votsch L4006 protocol. If you have these two chambers, you only need one MTCI unit. However, if you have another chamber like Espec SU-241 that uses […]

Kann die Arbin-Software einen Graphen erstellen, der Daten über Zyklen für einen oder mehrere Kanäle hinweg vergleicht?

Von Arbin Team | Mai 6, 2026

Yes. Arbin Data Watcher software (Version 2.2.217 and later) can do this. In Data Watcher software, click “Edit Chart Template”. From the Model dropdown, select “CycleCompare”. Return to the main window of Data Watcher. In the axis configuration, you can select the physical quantity for the X, Y1, and Y2 […]

Schrittsteuerungsfehlerprüfung unsicher

Von Arbin Team | Mai 6, 2026

The “Step Control Error Check Unsafe” warning indicates that the system has detected a significant deviation between the expected (configured) output and the actual measured output during a test step. They are designed to detect open circuits or short circuits caused by loose IV cable connections, it also detects abnormal […]

Mein Lade-/Entladestrom ist fast 0 oder viel niedriger als der eingestellte Wert?

Von Toan Nguyen | Mai 3, 2026

In some cases, the charge or discharge current may be significantly lower than the configured value, or even close to zero. This issue is typically caused by software configuration limits or hardware-related problems. This article provides a structured troubleshooting guide to help identify and resolve the issue efficiently. Symptoms You […]

Solar & Storage Live Malaysia 2026: Förderung von Batteriinnovationen für eine nachhaltige Energiezukunft

Von Arbin Team | April 12, 2026

Event Overview Solar & Storage Live Malaysia 2026 is a premier event in Southeast Asia dedicated to solar energy and energy storage technologies. Taking place on April 8–9, 2026 at MITEC in Kuala Lumpur, the exhibition gathers industry leaders, engineers, and innovators across the global energy ecosystem. Organized by Terrapinn, […]

Mein Test wurde aufgrund eines Fehlers namens “Tokenbroken” abgebrochen.”

Von Arbin Team | April 9, 2026

This error indicates too much data logged at a very short period of time. Check the schedule, make sure the data log setting is appropriate. Please open the schedule and locate the Log Limit: Then check the following points: Logging interval (time between data points)Many Arbin machines can acquire data […]

Warum sehe ich den Hilfswert nicht, obwohl ich ihn bereits zugeordnet habe?

Von Arbin Team | April 5, 2026

In some cases, users may successfully configure (map) an auxiliary channel, but the corresponding auxiliary values do not appear during test execution. This issue is typically caused by missing schedule assignment or incomplete auxiliary channel configuration within the schedule. Possible Causes No schedule or an incorrect schedule is assigned to […]

Warum stoppt mein Zeitplan sofort nach dem Start?

Von Arbin Team | April 2, 2026

In some situations, a schedule may stop immediately after it starts. This behavior usually occurs when certain safety conditions are not met or when required devices are not properly connected. The system automatically stops the schedule to prevent unexpected behavior, hardware damage, or unsafe operating conditions. This article provides common […]