Kann ich eine MTCI-Einheit kaufen und zwischen den Kammern wechseln?

Von Arbin Team | Mai 6, 2026

The answer is yes and no. Yes, if those chambers use the same communication protocol. For example, Votsch VT4002 and Weiss WT20/40 use the Votsch L4006 protocol. If you have these two chambers, you only need one MTCI unit. However, if you have another chamber like Espec SU-241 that uses […]

Kann die Arbin-Software einen Graphen erstellen, der Daten über Zyklen für einen oder mehrere Kanäle hinweg vergleicht?

Von Arbin Team | Mai 6, 2026

Yes. Arbin Data Watcher software (Version 2.2.217 and later) can do this. In Data Watcher software, click “Edit Chart Template”. From the Model dropdown, select “CycleCompare”. Return to the main window of Data Watcher. In the axis configuration, you can select the physical quantity for the X, Y1, and Y2 […]

Schrittsteuerungsfehlerprüfung unsicher

Von Arbin Team | Mai 6, 2026

The “Step Control Error Check Unsafe” warning indicates that the system has detected a significant deviation between the expected (configured) output and the actual measured output during a test step. They are designed to detect open circuits or short circuits caused by loose IV cable connections, it also detects abnormal […]

Mein Lade-/Entladestrom ist fast 0 oder viel niedriger als der eingestellte Wert?

Von Toan Nguyen | Mai 3, 2026

In some cases, the charge or discharge current may be significantly lower than the configured value, or even close to zero. This issue is typically caused by software configuration limits or hardware-related problems. This article provides a structured troubleshooting guide to help identify and resolve the issue efficiently. Symptoms You […]

Solar & Storage Live Malaysia 2026: Förderung von Batteriinnovationen für eine nachhaltige Energiezukunft

Von Arbin Team | April 12, 2026

Event Overview Solar & Storage Live Malaysia 2026 is a premier event in Southeast Asia dedicated to solar energy and energy storage technologies. Taking place on April 8–9, 2026 at MITEC in Kuala Lumpur, the exhibition gathers industry leaders, engineers, and innovators across the global energy ecosystem. Organized by Terrapinn, […]

Mein Test wurde aufgrund eines Fehlers namens “Tokenbroken” abgebrochen.”

Von Arbin Team | April 9, 2026

This error indicates too much data logged at a very short period of time. Check the schedule, make sure the data log setting is appropriate. Please open the schedule and locate the Log Limit: Then check the following points: Logging interval (time between data points)Many Arbin machines can acquire data […]

Warum sehe ich den Hilfswert nicht, obwohl ich ihn bereits zugeordnet habe?

Von Arbin Team | April 5, 2026

In some cases, users may successfully configure (map) an auxiliary channel, but the corresponding auxiliary values do not appear during test execution. This issue is typically caused by missing schedule assignment or incomplete auxiliary channel configuration within the schedule. Possible Causes No schedule or an incorrect schedule is assigned to […]

Warum stoppt mein Zeitplan sofort nach dem Start?

Von Arbin Team | April 2, 2026

In some situations, a schedule may stop immediately after it starts. This behavior usually occurs when certain safety conditions are not met or when required devices are not properly connected. The system automatically stops the schedule to prevent unexpected behavior, hardware damage, or unsafe operating conditions. This article provides common […]

Arbin nimmt als Firmensponsor am International Battery Seminar & Exhibit 2026 teil

Von Arbin Team | März 31, 2026

Arbin Instruments recently participated in the International Battery Seminar and Exhibit 2026 in Orlando, Florida—one of the industry’s most established events focused on battery technology and energy storage innovation. As a proud Corporate-level sponsor, Arbin was honored to support this premier gathering of researchers, engineers, and industry leaders from across […]

Arbin präsentiert fortschrittliche Batterietestlösungen auf der The Battery Show Asia 2026 in Hongkong

Von Arbin Team | März 13, 2026

Arbin Instruments proudly participated in The Battery Show Asia 2026, one of the region’s most influential exhibitions for advanced battery technology, energy storage, and electric mobility. The event was held at AsiaWorld‑Expo in Hong Kong, bringing together leading companies, engineers, and innovators from across the global battery ecosystem. As both […]

Arbin präsentiert fortschrittliche Batterietestlösungen auf der InterBattery 2026 in Seoul

Von Arbin Team | März 13, 2026

Arbin Instruments is pleased to announce its participation in InterBattery 2026, one of the world’s leading exhibitions dedicated to battery technology and energy storage innovation. The event takes place March 11–13, 2026, at COEX Convention & Exhibition Center in Seoul, South Korea. As Korea’s largest international battery exhibition, InterBattery brings […]

Anwendungsbeispiele für benutzerdefinierte MetaVariablen (MV_UD)

Von Arbin Team | 26. Februar 2026

In Arbin software, MV_UD (short for User-Defined Meta Variables) is a powerful tool for controlling battery testers. This article explains how to use MV_UD in Arbin MITS software. Storing Values in MV_UD with SetValue(s) In the Control Type list, select SetValue(s). SetValue(s) assigns a fixed value of a physical quantity […]

So plottest du ein Potenzial-gegen-Strom-Diagramm mit dem Arbin Data Watcher

Von Arbin Team | 4. Februar 2026

1. Introduction 1.1 Background Arbin DataWatcher is primarily used as a real-time data analysis and visualization tool. It allows users to monitor test data, plot curves, and export data files efficiently. It enables users to visualize and analyze key electrochemical parameters such as voltage, current, capacity, and time. 1.2 Purpose […]

Genauigkeit und Präzision: Auswirkung der Präzision

Von Arbin Team | 22. Januar 2026

We previously examined the impact of measurement accuracy in an earlier article. In this article, we shift our focus to the influence of a battery tester’s precision on battery research. In many research applications, parameters such as dV/dt and dQ/dV are widely used. The reliability of these measurements depends heavily […]

Arbin Instruments at The Battery Show Asia 2026

Von Arbin Team | Januar 12, 2026

We are excited to share that Arbin Instruments will be participating in The Battery Show Asia 2026, held from March 10–12, 2026 at AsiaWorld-Expo in Hong Kong — one of the most influential gatherings in the global battery and energy storage industry. This premier three-day business-to-business exhibition and conference brings […]

Genauigkeit und Präzision: Auswirkungen der Genauigkeit des Batterietesters

Von Arbin Team | 8. Januar 2026

In the previous article, we introduced the concepts of accuracy and precision, along with a brief overview of their influence on battery testing. In this article, we will take a deeper look at the impact of accuracy. We will take an example about the impact on capacity measurement and Coulombic […]

Genauigkeit und Präzision: Überblick

Von Arbin Team | 23. Dezember 2025

1. Introduction Battery testing is fundamentally a measurement task. Whether evaluating cycle life, coulombic efficiency, voltage response, or degradation mechanisms, the quality of conclusions depends on the quality of the data. Two concepts govern data integrity in any battery test system: accuracy and precision. Although often mentioned together, they describe […]

Übersicht, welches Arbin-Produkt Sie wählen sollen – Regenerative Funktionen

Von Arbin Team | 27. November 2025

Energy efficiency is a crucial concern for every factory and department. Arbin’s regenerative products are designed to meet your energy-saving needs with exceptional performance and reliability. I. Understanding the Regenerative Feature The regenerative feature allows energy released during battery discharge to be returned to the power grid, instead of being […]

Testen von EV-Batteriepacks unter realen Fahrbedingungen mithilfe von Simulationssteuerungstypen

Von Arbin Team | 5. November 2025

Information and Instructions for using the Simulation control types. 1. General Description Simulation refers to the ability to input data collected from a non-formulated dynamic regime (i.e. other than Constant Current, Constant Voltage, Constant Power, Ramp, Staircase, Pulse, or Formula) as a control function. There are three control types: 1. […]

Führungsbatterie-Innovation: Arbin auf der Battery Show India 2025

Von Arbin Team | 22. Oktober 2025

We are excited to share that Arbin will be participating in The Battery Show India 2025, taking place from October 30 to November 1 at the India Expo Mart & Centre in Hall 7, Booth B352 , Greater Noida. As a global leader in high-precision battery testing technology, Arbin provides […]