Are there specific requirements for preparing the facility to set up the Arbin system?

Generally, there are several points to note:

  1. Arbin machines typically have a working ambient temperature range of 0°C to 30°C. Therefore, please prepare the air-conditioning system accordingly.

  2. The power grid system must be prepared to match the machine’s input power requirements. For example, some machines require single-phase power input, while others require a three-phase power supply.

  3. You also need to pay attention to the input current requirements. Please make sure the power socket can provide sufficient current for either the single-phase or three-phase input of the Arbin machine.

  4. The distance between the machine and the wall should also be considered:

    Linear system Layout Recommendation
    (Applied for  LBT, LBTS, HPS, HS series)

    The distance from the right wall should be ≥ 0.5m,
    The distance from the left wall should be ≥1.5m,
    The distance from the back wall should be ≥1m.

arbin linear systems layout

Linear systems’ layout

RBT system Layout Recommendation
(Applied for RBT3, RBT4 series)

The distance from the right wall should be ≥ 1.5m
The distance from the left wall should be ≥1.5m
The distance from the front and back wall should be ≥2m.

arbin rbt systems layout

RBT systems’ layout

If you have more than one Arbin system, please contact your sales rep to get more information about the distance between the systems and layout recommendation.