Generally, there are several points to note:
-
Arbin machines typically have a working ambient temperature range of 0°C to 30°C. Therefore, please prepare the air-conditioning system accordingly.
-
The power grid system must be prepared to match the machine’s input power requirements. For example, some machines require single-phase power input, while others require a three-phase power supply.
-
You also need to pay attention to the input current requirements. Please make sure the power socket can provide sufficient current for either the single-phase or three-phase input of the Arbin machine.
-
The distance between the machine and the wall should also be considered:
Linear system Layout Recommendation
(Applied for LBT, LBTS, HPS, HS series)The distance from the right wall should be ≥ 0.5m,
The distance from the left wall should be ≥1.5m,
The distance from the back wall should be ≥1m.
RBT system Layout Recommendation
(Applied for RBT3, RBT4 series)
The distance from the right wall should be ≥ 1.5m
The distance from the left wall should be ≥1.5m
The distance from the front and back wall should be ≥2m.
If you have more than one Arbin system, please contact your sales rep to get more information about the distance between the systems and layout recommendation.