올해는 배터리 및 에너지 저장 테스트 분야의 선두주자인 Arbin Instruments가 30주년을 맞이하는 중요한 해입니다. 배터리 테스트 요구 사항에 대해 Arbin이 어떻게 도움을 줄 수 있는지 자세히 알아보세요.
블로그
주제
가장 인기 있는
최신
Solar & Storage Live Malaysia 2026: Advancing Battery Innovation for a Sustainable Energy Future
Event Overview Solar & Storage Live Malaysia 2026 is a premier event in Southeast Asia dedicated to solar energy and energy storage technologies. Taking place on April 8–9, 2026 at MITEC in Kuala Lumpur, the exhibition gathers industry leaders, engineers, and innovators across the global energy ecosystem. Organized by Terrapinn, a leading international B2B event […]
My test stopped due to an error called “Tokenbroken”
This error indicates too much data logged at a very short period of time. Check the schedule, make sure the data log setting is appropriate. Please open the schedule and locate the Log Limit: Then check the following points: Logging interval (time between data points)Many Arbin machines can acquire data at rates of up to […]
Why can’t I see the auxiliary value even though I have already mapped it?
In some cases, users may successfully configure (map) an auxiliary channel, but the corresponding auxiliary values do not appear during test execution. This issue is typically caused by missing schedule assignment or incomplete auxiliary channel configuration within the schedule. Possible Causes No schedule or an incorrect schedule is assigned to the channel Forgot to add […]