Conflict Clamp V” pop up when I start a test

This error occurs only when the 基于通道的电压钳位--- option is unchecked in ArbinAdvSys.cfg内可以. This means all channels on the same channel board share one voltage clamp/voltage check. The MITS software must ensure all schedules on the same board channels share the same Voltage Clamp setting; otherwise, the Conflict Clamp V error appears.

最近,几乎没有Arbin系统共享电压钳,所以大多数情况下,这种不安全是由于错误的高级功能设置。只需检查 基于通道的电压钳位---ArbinAdvSys.cfg内可以  解决这个问题。

arbinadvsys

arbin system setup

Please note you may need an Arbin support engineer’s help to make this change because it requires an sa account, not an admin account.